Warning: Each time changes are made to the circuit configuration, confirm that delivered pressures are
adequate for patient management.
Warning: The addition of accessories to the patient circuit (e.g., heated humidifier, bacteria filter, water
traps, and additional tube lengths) affects pressure and flow delivery at the patient
connection. Clinical assessment must confirm that the delivered pressures and flows are
adequate for patient management when circuit accessories are used. It may be necessary to
increase pressure delivery to compensate for the increased resistance of the circuit.
Setting Up the BiPAP S/T System
Note: Heat and moisture exchangers should not be used with the BiPAP S/T System. Testing has
shown that the ability of the unit to maintain the prescribed pressure is significantly affected
by the addition of a heat and moisture exchanger to the patient circuit.
3. See Figure 18. Plug the electrical cord into the power
connector on the back of the unit.
4. See Figure 19. Make sure the voltage selector switch on the
back of the unit is set correctly.
United States and Canada setting: 115 V
European and other countries: Check the country’s line
voltage requirements for the correct setting.
5. See Figure 20. Make sure the ON/OFF switch is OFF (0), then plug
the electrical cord into a three-prong outlet.
Note: In the United States and Canada, the use of a three-prong to
two-prong adapter or extension cord is not recommended.
6. Turn the unit on, occlude the outlet port, and confirm the presence
of flow from the exhalation port.
115
230
Power Cord
Connector
Figure 18
115
230
115
230
Figure 19
Figure 20