40
Specifications
Oxygen Saturation Display Range 0 to 100% SpO
2
Pulse Rate Display Range 18 to 321 beats per minute (BPM)
Saturation Declared Accuracy (A
rms
)*
70-100%
No Motion Adults / Pediatrics Neonates
Finger Clip: ±2 digits N/A
Flex, FlexiForm: ±3 digits ±4 digits
Ear Clip: ±4 digits N/A
Motion
Finger Clip: ±3 digits N/A
Low Perfusion
Finger Clip: ±3 digits N/A
Flex, FlexiForm: ±3 digits ±4 digits
Pulse Rate Declared Accuracy 18 to 300 BPM (no motion)
40 to 240 BPM (motion)
40 to 240 BPM (low perfusion)
No Motion Adults / Pediatrics Neonates
Finger Clip:
±3 digits N/A
Flex, FlexiForm: ±3 digits ±3 digits
Ear Clip: ±3 digits N/A
Motion
Finger Clip: ±5 digits N/A
Low Perfusion
Finger Clip: ±3 digits N/A
Finger Clip, Flex, FlexiForm: ±3 digits ±3 digits
Measurement Wavelengths and Output Power**
Red: 660 nanometers @ 0.8 mW max. avg.
Infrared: 910 nanometers @ 1.2 mW max. avg.
Sensor Families:
936, 936P
Flex Sensors: 953, 934, 935
FlexiForm Sensors: 953D, 937D, 934D, 935D
* ± 1 A
rms
represents approximately 68% of measurements.
** This information is especially useful for clinicians performing photodynamic therapy.