User ManualautoSVTMwithEncore SmartCard®
6User ManualIPAPPressureEPAPPressureRise TimeTimeFigure 3–1 IPAP and EPAP Breathing LevelsYou can adjust the Rise Time to make the pressure change mo
User Manual7NOTE: e exhalation port may be part of the mask or may be part of a separate exhalation device, but is required to minimize the potentia
8User ManualChapTer 4: deviCe ConTrols and display feaTuresFigure 4–1 shows the location of the device’s alarm power indicators, control panel, Press
User Manual94.2 Control Panele control panel contains the following control buttons and indicators.4.2.1 Control Buttonse control buttons on the c
10User Manual4.2.2 alarm and PoWer indiCatorsFigure 4–4 shows the device’s alarm and power indicators.AC PowerIndicator (Green)DC PowerIndicator (Gre
User Manual11e information shown on the display screen is defined as follows:ALARMIndicates that the device requires user attention as indicated on th
12User Manual4.2.4 Breathing CirCuit ConneCtionFigure 4–6 shows where the circuit tubing connects to the device.BreathingCircuit ConnectionCircuitTub
User Manual13ChapTer 5: seTTing up The bipap auTosv deviCe5.1 installing the air filterse device uses one or two removable filters at the air inlet.
14User Manual5.2 Where to PlaCe the deviCePlace the device on its base somewhere within easy reach of where you will use it. Make sure that the air i
User Manual15B. If you are using a mask with a separate exhalation device, connect the open end of the circuit tubing to the exhalation device as sho
© 2007 Respironics, Inc. and its afliates. All rights reserved.This BiPAP system is covered by one or more of the following patents: U.S. Patent Nos.
16User Manual5.4 ComPlete setuPFigure 5–6 shows the completed breathing circuit setup.BreathingCircuit ConnectionCircuitTubingPatient InterfaceBacter
User Manual173. Leaving a small amount of slack in the cord, connect the cord on the other side of the power supply to one of the power inlets on the
18User ManualChapTer 6: operaTing The deviCe6.1 starting the deviCe1. Plug in the device to an AC or DC power source to power up the device. A confi
User Manual19NOTE: With the exception of the button, the control pad is inactive during these first three screens. Each of these screens appears for
20User Manual6.2 Changing the deviCe settingsYou can view the following settings and indicators on the display screen:• Measured pressure• Backligh
User Manual213. You can exit this screen by pressing the Left or Right User buttons or the SILENCE button. For additional information on using a hum
22User Manual1. From either the Monitoring or Standby screens, press the Right User button until you reach this screen. e Rise Time Setting screen i
User Manual231. From either the Monitoring or Standby screens, press the Right User button until the LED Backlight Setting screen appears, as shown i
24User Manual1. Leak Screenis screen, shown in Figure 6–12, shows the average of the leak values for the previous six breaths.LEAKFigure 6–12 Leak
User Manual25ChapTer 7: alarms7.1 introduCtion to alarmse device provides three alarm levels: high, medium, and low priority.High Priority ese alar
Table of ConTenTsC : P C ...
26User Manual7.2 What to do When an alarm oCCurse following example applies to most alarm conditions. Follow these steps unless otherwise directed b
User Manual277.3 alarm taBlese following tables summarize the high priority, medium priority, and low priority alarms. 7.3.1 high Priority alarmsAl
28User Manual7.3.2 medium Priority alarmsAlarmLEDAlarmSoundDisplayMessageDeviceActionPossibleCauseYour ActionYellow FlashDC PowerLED Flashes• • •Oper
User Manual29ChapTer 8: TroubleshooTingis chapter describes problems that you may experience with your BiPAP autoSV device or mask and provides possi
30User ManualProblem Why It Happened What To DoRedness occurs when the mask cushion comes in contact with the skin.is could be due to improper mask fi
User Manual31Problem Why It Happened What To Doe device’s display is erratic.e device or power supply has been dropped or mishandled, or the device
32User ManualChapTer 9: Cleaning and mainTenanCe9.1 Cleaning the deviCeBefore cleaning or performing any routine maintenance, always make sure the d
User Manual333. Remove the filters from the enclosure by gently pulling around the edges of the filters. e top filter is the reusable gray foam filter.
34User ManualChapTer 10: aCCessories10.1 adding a humidifiere REMstar Heated Humidifier, REMstar Passover Humidifier, and H2 Heated Humidifier are a
User Manual35ChapTer 11: speCifiCaTionsenvironmentalOperating StorageTemperature41° F (5° C) to 95° F (35° C)-4° F (-20° C) to 140° F
C : A ...
36User ManualPressureOutput: 4 to 30 cm H2OControl aCCuraCyParameter Range AccuracyIPAP Min 4 to 30 cm H2O ± 5 cm H2O*IPAP Max 4 to 30 cm H2O ± 5 cm H
User Manual37appendix a: emC informaTionguidanCe and manufaCTurer’s deClaraTion - eleCTromagneTiC emissions: is device is intended for use in the el
38User ManualguidanCe and manufaCTurer’s deClaraTion - eleCTromagneTiC immuniTy: is device is intended for use in the electromagnetic environment spe
limiTed WarranTyRespironics, Inc. warrants that the BiPAP autoSV system shall be free from defects of workmanship and materials and will perform in ac
1040200JR 3/1/071001 Murry Ridge Lane Respironics DeutschlandMurrysville, PA 15668 Gewerbestrasse 17USA 82211 Herrsching, Germany
User Manual1ChapTer 1: paCkage ConTenTsYour BiPAP autoSV device should include the following items. If any of these items are missing, contact your he
2User ManualChapTer 2: Warnings and CauTionsWARNING: Indicates the possibility of injury to the user or operator.CAUTION: Indicates the possibilit
User Manual3 Explanation of the Warning: e device is intended to be used with special masks or connectors that have exhalation ports to allow contin
4User Manual2.2 Cautions• e device may only be operated at temperatures between 41° F (5° C) and 95° F (35° C).• A properly installed, undamaged r
User Manual5ChapTer 3: inTroduCTion To The deviCe3.1 definitionse following terms appear throughout this manual:Apnea A condition marked by the cess
Comments to this Manuals