38
7.3.2 MEDIUM PRIORITY ALARMS
Alarm
LED
Alarm
Sound
Display
Message
Harmony
Action
Possible
Cause
Your
Action
Yellow Flash
• • •
DC Power
LED
Flashes
Operates Battery nearly
discharged.
Replace the battery.
Alarm
LED
Alarm
Sound
Display
Message
Harmony
Action
Possible
Cause
Your
Action
Yellow Solid
• •
CARD
flashes and
card error
code (Cxx)
displays
Operates
There is a problem
with the SmartCard
inserted in the
SmartCard
connectivity slot.
Remove the SmartCard
from the device and
contact your home
care provider.
Yellow Solid
• •
ALARM,
CARD, and
cmH
2
O
flash and
LCD light
is on
Operates
The SmartCard
storage capacity
is reached.
DC power
LED flashes
Blank screen
Operates
Harmony lost AC
power and is now
operating on DC
power.
Check AC power and
seek reliable power
source.
At start-up only,
alarm notifies you
that a battery is
being used
to provide power.
Provide AC power if
you do not want to use
a battery; otherwise,
no action needed.
Remove the SmartCard
from the device and
contact your home
care provider.
Yellow Solid
• •
7.3.3 LOW PRIORITY ALARMS