CHAPTER 11: SPECIFICATIONS
ENVIRONMENTAL
Operating Storage
Te mperature 5ºC to 35 ºC -20ºC to 60ºC
Relative Humidity 15 to 95% 15 to 95%
(non-condensing) (non-condensing)
Atmospheric Pressure 83 to 102kPa
(5600 feet to sea level)
PHYSICAL
Dimensions: 9.75” L x 6.625” W x
4.4” H
Weight: Approximately 4.1 lb
ELECTRICAL
AC Voltage: 100 to 240 V, 50/60 Hz
DC Voltage: 12 V (when operated
with the external DC
power supply)
AC Current: 1.25 A maximum
DC Current: 3.0 A maximum
Protection against electric shock: Class II
Degree of protection against electric shock: Type BF Applied Part
Degree of protection against
harmful ingress of water: Harmony unit:
Ordinary Equipment,
IPX0
AC Power Supply:
Ordinary Equipment,
IPX1
DC Boost Converter:
Ordinary Equipment,
IPX1