Page view 14
Basic Operation 9
Power On/Standby
Turn on the 920M PLUS by pressing and releasing the On/Standby
switch . Turn off the unit by pressing and holding this switch for about 2
seconds. Refer to Figure 2 on page 4 for a detailed illustration of the switches
and indicators. Refer to “Detailed Operation” on page 11 for detailed
information about entering setup mode and on general operation of the
920M PLUS.
Power On Self-Test
When the 920M PLUS is powered on for normal operation by pressing and
releasing the On/Standby switch , the unit will cycle through a startup (or
initialization) sequence before displaying valid data values.
During a startup, always check for any missing indicators or LED display
segments. If any indicator is not lit, do not use the 920M PLUS. Contact
Respironics Customer Support for repair or replacement.
At normal startup (when not entering the setup mode), with fresh batteries,
the 920M PLUS will cycle through the following sequence:
will appear briefly in the SpO
2
and pulse rate displays
the yellow low battery indicator will turn on steadily for a few seconds
the pulse quality indicator will turn on red for 1 second, then green for 1
second, and then begin to blink
the clock time currently set in the memory (in hours and minutes, for
example) appears briefly in the displays
the software revision number (the letter “r” followed by a 3-digit number,
for example) appears briefly in the displays
(two dashes) will appear in the displays until a valid pulse signal is
detected
Note: When entering setup mode at power on (for memory clear, calendar, or
clock settings), the 920M PLUS will cycle through a startup sequence
similar to the above, with the following exceptions: 1) The pulse quality
indicator will first turn on red for 1 second, then green for 1 second, but
will then turn off and not blink until after exiting setup mode; 2) the
software revision number will not appear; and 3) the two dashes will
appear only after exiting setup mode.
888
888
04
41
r
018
-
-