Detailed Operation 15
Displays and Indicators
SpO
2
Display
The SpO
2
display is the upper numeric display. (Identified by the %SpO
2
symbol; see Figure 2 on page 4.) This 3-digit light-emitting diode (LED)
display shows the current oxygen saturation percentage.
Pulse Rate Display
The pulse rate display is the lower numeric display. (Identified by the ♥
symbol; see Figure 2 on page 4.) This 3-digit LED display shows the pulse
rate in pulses per minute.
Pulse Quality Indicator
The pulse quality indicator, a tricolor LED, blinks once for each detected
pulse. (Identified by the symbol; see Figure 2 on page 4.) The pulse quality
indicator changes color to indicate changes in the pulse waveform signal that
may affect the SpO
2
data.
The pulse quality indicator may blink one of three colors: green, yellow, or
red. These colors are similar to the colors of a stoplight:
• Green indicates that the pulse waveform signal is of good quality and the
SpO
2
and pulse rate data are accurate.
• Yellow indicates that the pulse waveform amplitude is marginal or that the
pulse oximeter has detected artifact. Although the SpO
2
and pulse rate data
may be acceptable, corrective measures should be considered if the
indicator continues to blink yellow frequently. To improve the signal
quality, try repositioning the sensor, try a different sensor type, eliminate
patient movement, or improve circulation at the site by massaging the area.
• Red indicates that the pulse waveform amplitude is inadequate. While the
pulse quality indicator is red, SpO
2
and pulse rate values are frozen (not
updated). After approximately 20 seconds, the values are replaced with
dashes indicating that SpO
2
and pulse rate measurements are not possible.
CAUTION: The 920M PLUS might misinterpret motion as good
pulse quality. Minimize patient motion as much as
possible.
2
Comments to this Manuals